Bien Airotor - Prestige L
Airotor handpiece
- Type - Turbine with light, small head
- Connection - Using 4-way Unifix quick-connect coupling, rotating 360 degree
- Speed of rotation - 335,000 rpm
- Recommended air pressure - 3.1 bar
- Cooling - Diffuser, with 3 separate air and water sprays, water antiretraction valve
- Chuck locking mechanism - Push-button with anti-heating system
- Light transmission - Multi-strand optical glass conductor
- Lighting - Halogen
- Cleaning/Sterilisation - Thermo washer disinfectable / Steam autoclave up to 135 degree C
- Weight - 73 g
- Prestige L ref. - 1600380-001
- Prestige L + 4-way Unifix set ref. - 1700184-001
MAINTENANCE OF TURBINES.
CLEANING
B.
REMOVING DIRT/DEPOSITS
- Carefully remove all traces of dirt or deposits from the spray nozzles using the Bien-Air cleaning wire
- Dry the nozzles with compressed air then dry the rest of the instrument with a cloth
Do not immerse in an ultrasonic cleaner
C.
INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET
- Shake the Spraynet before use
- Use the Spraynet attachment 1
(for 2-/3-way and 4-way turbines, see inset below) - Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
- Clean the exterior of the turbine with a cloth soaked in Spraynet
FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
C'.
CLEANING THE INSIDE WITH SPRAYNET FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
- Use the Spraynet attachment 2
- Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second
Spray into the opening as shown in the illustrations for “2-/3-way” and “4-way”
DISINFECTION
OR
A.
DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer's instructions for use)
Do not immerse the instrument in a disinfectant bath
LUBRICATION
OR
B.
LUBRICATION WITH LUBRIFLUID
- Shake the Lubrifluid before use
- Use the Lubrifluid attachment 2, 3 ou 4depending on the type of instrument * (For 2-/3-way and 4-way turbines, see inset below)
- Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
- Then, run an operating test for 30 seconds at low speed to remove any surplus lubricant
- Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out
FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
B'.
INTERNAL LUBRICATION WITH LUBRIFLUID FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
- Use the Lubrifluid attachment 2
- Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second
Spray into the opening as shown in the illustrations for “2-/3-way” and “4-way”